Biological Safety

Laboratory Solutions offers Biological Safety Cabinets manufactured by Labconco Corporation, USA. Labconco is a ISO 9001 certified company that manufactures products that adhere to high standards including ETL and NSF standard 49.

Purifier Logic®+ Class II, Type A2 and Type B2 Biological Safety Cabinets provide personnel, product and environmental protection from hazardous particulates and agents that require Biosafety Level 1, 2 or 3 containment. Purifier Cell Logic+ Type A2 and B2 Biosafety Cabinets have features specialized for cell research.

The Purifier Axiom® Class II, Type C1 is the first biological safety cabinet that can be used in a recirculating Type A mode for standard microbiological work, or can be connected to an exhaust system to function in Type B mode for handling hazardous chemical vapors or radionuclides. The Axiom is a finalist for the 2016 R&D 100 Awards for Innovation.

Purifier Class I and HEPA Filtered Enclosures provide personnel and environmental protection and are ideal for procedures that generate fine dust, handling suspicious mail, or weighing hazardous powders. They offer an economical alternative to Class II cabinets when applications do not require product protection.

Purifier Cell Logic+ Class II, Type A2 Biosafety Cabinets have
unique features that address the unique needs of cell research.
Some models have a modified sash that accommodates a
microscope. Other models have a temperature-controlled work
area on the work surface to keep cells at the proper temperature
for viability.
Both Purifier Logic+ and Cell Logic+ Class II, Type A2 Biosafety
Cabinets have the same airflow pattern to achieve personnel and
product protection. During operation, room air is drawn into the
inlet grille at the work access opening and through a supply HEPA
filter. Approximately 70% of the HEPA-filtered air is circulated
through the cabinet, while 30% passes through an exhaust HEPA
filter and is discharged. These biohazard cabinets are designed to
discharge HEPA-filtered exhaust air directly into the laboratory, or
into an exhaust system through the optional Canopy Connection.
When canopy- ducted, these biological safety cabinets may be
used for applications involving minute quantities of volatile toxic
chemicals and tracer amounts of radionuclides as an adjunct to
microbiological research.

Purifier Cell Logic+ Class II, Type B2 Biosafety Cabinets
have unique features that address the unique needs of cell
research. Some models have a modified sa sh that
accommodates a microscope. Other models have a
temperature-controlled work area on the work surface to
keep cells at the proper tempera ture for viability.
Both Purifier Logic+ and Cell Logic+ Class II, Type B2
Biosafety Cabinets have the same airflow pattern to
achieve personnel and product protection.
These biological safety cabinets are designed to be hard
ducted to the outside. During operation, room air is drawn
into the top of the cabinet and through a supply HEPA
filter. This filtered air flows downward through the work
area. Room air is also drawn into the inlet grille located at
the work access opening. All of the contaminated air
passes through the exhaust HEPA filter. A dedicated
exhaust system and remote blower discharge 100% of the
filtered exhaust air from the laboratory. Since none of the
air is recirculated, these biohazard cabinets may be used
for work with agents treated with volatile toxic chemicals
and radionuclides.

The Purifier Axiom Class II, Type C1 is the first biological
safety cabinet that can be used in a recirculating Type A
mode for standard microbiological work, or can be
connected to an exhaust system to function in Type B
mode for handling hazardous chemical vapors or
radionuclides. Because of the efficient, well-defined
airflow pattern over its work surface, the Axiom is safer
and easier to operate than traditional biosafety cabinets.
The patent-pending design of the Chem-Zone TM
maximizes chemical safety by creating a large, clearly
defined area within the enclosure, bringing peace of mind
during chemical handling. Other features that make the
Axiom safe and easy to operate include its ergonomic
InclinationTM Technology, the intuitive MyLogicTM
Operating System and Active Protection Protocol.
The Axiom provides protection for personnel, product and
the laboratory environment by filtering hazardous
particulates and aerosols, such as agents that require
Biosafety Level 1, 2 or 3 containment, with the option of
venting away hazardous fumes. Type C1 enclosures may
also be used for a wide va riety of applications that
incorporate antineoplastic drugs, genetic material,
carcinogens, allergens, or other airborne hazards.